Flow Cytometer

XF-1600

Instructions for Use

for upgraded XF-1600 to

14 fluorescence detection channels

Code No. CF819831 en-eu

PRINTED IN JAPAN

Date of Issue or Revision: 04/2025

Document Version: 1.0

Software Version: 0 onwards

© SYSMEX CORPORATION

Revision history

XF-1600 Instructions for Use

3/14

4

Revision history

04/2025

Initial issue

Document version: 1.0

Software version: 0

Revision history

XF-1600 Instructions for Use

4/14

Table of contents

XF-1600 Instructions for Use

5/14

6

Table of contents

Revision history 3

Chapter 1 Introduction 7

1.1 Points to note about this manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

1.2 Symbols used in this manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

Chapter 2 Instrument specifications 9

2.1 Optics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

Chapter 3 Cytometer QC and setup 11

3.1 Registering lot information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

Chapter 4 Registering panel/subset panel and

workitems 13

4.1 Reviewing subset panels and QC files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13

4.2 Workitems. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13

Table of contents

XF-1600 Instructions for Use

6/14

Chapter 1 Introduction

XF-1600 Instructions for Use

7/14

Chapter 1 Introduction

Thank you for purchasing the 14 COLOR UPGRADE KIT for XF-1600. This manual explains how to operate

the Sysmex Flow Cytometer XF-1600 when upgraded to 14 fluorescence detection channels by using the 14

COLOR UPGRADE KIT.

Read and understand the contents of this manual before operating this product, and keep this manual in a

safe place for future reference.

This is not an inclusive manual. The explanations in this manual assume that you have already read the

XF-1600 Instructions for Use.

For basic operations of the instrument, see the XF-1600 Instructions for Use.

In vitro diagnostic medical device

Contact Addresses

1-5-1 Wakinohama-Kaigandori, Chuo-ku, Kobe 651-0073, Japan

Authorized representatives

Europe, Middle East and Africa

Bornbarch 1, 22848 Norderstedt, Germany

Tel +49 40 5 27 26-0 / Fax +49 40 5 27 26-100

Asia-Pacific

9 Tampines Grande #06-18 Singapore 528735

Tel +65-6221-3629 / Fax +65-6221-3687

Service and maintenance

Please contact your authorized local Sysmex representative.

The system described in this manual is marked with the CE mark, which

confirms compliance with the applicable requirements of:

• Regulation (EU) 2017/746 on in vitro diagnostic medical devices

• Directive 2011/65/EU on the restriction of the use of certain hazardous

substances in electrical and electronic equipment and its related amending

directives including (EU) 2015/863 amending Annex II of Directive 2011/65/

EU

Note:

• Data generated by the XF-1600 is not intended to replace professional judgment in the determination of

a diagnosis or in monitoring patient therapy.

• Operate the instrument as instructed. Reliability of test results cannot be guaranteed if there are any

deviations from the instructions in this manual. If the instrument fails to function properly as a result of

either a user's operation not specified in this manual or a user's utilization of a program not specified by

Sysmex, the product warranty will be void.

Chapter 1 Introduction

XF-1600 Instructions for Use

8/14

1.1 Points to note about this manual

• You may not reproduce the contents of this manual in whole or in part without permission from Sysmex

Corporation.

• Images in this manual related to the product are for illustrative purposes only and may not exactly match the

product itself.

• While we have taken all possible precautions to ensure the quality of this manual, please contact your authorized

local Sysmex representative if you find any errors or omissions.

1.2 Symbols used in this manual

Note:

Indicates useful information for using the instrument.

Chapter 2 Instrument specifications

XF-1600 Instructions for Use

9/14

Chapter 2 Instrument specifications

2.1 Optics

The semiconductor laser light (488 nm, 638 nm, and 405 nm) interrogates the cells and particles passing through the

flow cell. The forward scattered light is detected by a photodiode, and the side-scattered light and fluorescence are

detected by a photomultiplier. The light signals received by each detector are converted into electrical signals, and by

analyzing these signals, cell and particle information can be obtained.

Channel Laser wavelength Fluorochromes or FSC/SSC Optical filters

FSC 488 nm Forward scatter -

SSC Side scatter -

FL1 FITC 530/30

FL2 PE 580/30

FL3 PE-DyLight 594 610/14

FL4 PerCP-Cy5.5 700/50

FL5 PE-Cy7 750LP

FL6 638 nm APC 660/20

FL7 Alexa Fluor 700 720/40

FL8 APC-Cy7 780/60

Laser Diode (488 nm)

Flow cell

FL6 FL7

FL2 FL1

FL4 FL3

FL9

FL10

Laser Diode (638 nm) Laser Diode (405 nm)

FL8

Forward Scatter (FSC)

Side Scatter (SSC)

FL5

FL11

FL12

FL13

FL14

Chapter 2 Instrument specifications

XF-1600 Instructions for Use

10/14

* For information about the fluorochromes that are suitable for your use environment, please contact your

authorized local Sysmex representative.

FL9 405 nm Pacific Blue 450/50

FL10 Pacific Orange 540/80

FL11 -* 603/30

FL12 -* 670/24

FL13 -* 710/50

FL14 -* 750LP

Channel Laser wavelength Fluorochromes or FSC/SSC Optical filters

Chapter 3 Cytometer QC and setup

XF-1600 Instructions for Use

11/14

Chapter 3 Cytometer QC and setup

3.1 Registering lot information

Once the instrument is upgraded to 14 fluoroescence detection channels, QC files that were set up before the

upgrade cannot be used for measurement. To use the same lot of the quality control material, a new QC file must be

created.

For details on how to create a new QC file, see the following section:

("Chapter 9: 9.7 Registering lot information" in the XF-1600 Instructions for Use)

Chapter 3 Cytometer QC and setup

XF-1600 Instructions for Use

12/14

Chapter 4 Registering panel/subset panel and workitems

XF-1600 Instructions for Use

13/14

14

Chapter 4 Registering panel/subset panel and workitems

4.1 Reviewing subset panels and QC files

When the instrument is upgraded to 14 fluorescence detection channels, the positions of populations may move

slightly. Perform the following procedure:

1 Review the subset panel analysis conditions that are registered before the

upgrade.

("Chapter 10: 10.3 Registering subset panel analysis conditions while running a sample" in the XF-1600

Instructions for Use)

2 Review the baseline of related panel QC files.

("Chapter 11: 11.9 Managing panel QC files" in the XF-1600 Instructions for Use)

4.2 Workitems

Note:

• The workitems that are measured before upgrading the instrument to 14 fluorescence detection

channels do not include data for the additional fluorescence detection channels.

• Do not measure workitems that are registered before the upgrade